Requirements, mounting, focusing and care for smartCam, proCam XS and scopeCam, and the full instructions for use to download.
What the computer, the cables and the light need for a sharp 4K image.
VISION 1.2 runs on Windows 10 or 11 (64-bit) and on Macs with Apple silicon. For 4K, plan on 16 GB of memory and a display of at least 1920 × 1080. You can record 4K even if your screen is not 4K. Full requirements
The camera and the foot pedal each need a USB port. Beyond 2 m, use the supplied active extension cables, up to 10 m in all. Longer runs can break up the image.
Room light of at least 10,000 lux and a treatment light of 30,000 lux or more, around 5,500 K, with a colour rendering index (CRI) of 95 or higher. Avoid halogen lights: they turn the image yellow.
Keep the treatment light and the room light close in brightness. A bright task light in a dark room gives a harsh, poorly focused image. Add room light, or move in closer to the bright area.
If you use electrosurgery, or the computer is not rated for medical use (IEC 60601-1), connect the camera through a USB isolator with 5 kVrms protection.
Turn off the computer's power saving while you record. The camera is powered through its USB cable and stops when the port sleeps.
Check the contents before you start. The cables, pedal and mounts are also sold as spare parts.
The camera in a zip case, with the ball-joint mount and the overhead light adhesive mount; the focus card; a 1.5 m black micro-USB camera cable; a Torx T6 screwdriver; the quick guide and the instructions for use.
The loupe attachment (installed on the camera), a universal loupe clip, a spiral cable guide for the loupe arm, a cable guide with a clothing clip, and a 1.5 m micro-USB to USB-C cable for Android devices. The 16 mm lens is installed; 8, 25 and 35 mm lenses are available.
The USB foot pedal (1.2 m cable), a 5 m white micro-USB camera cable, two 5 m active USB extension cables, 15 curved and 25 flat cable clips, and the universal overhead light attachment with screws.
scopeCam comes with its own adapters for the microscope. Its contents are listed in the scopeCam instructions for use.
Where each camera goes, step by step. Mount the camera before you focus it.
smartCam and proCam XS
1
Wipe the light's housing with isopropyl alcohol where the mount will go.
2
Centre the camera on the light so both point at the same treatment point. Keep clear of the light's sensors and its hot surfaces: the adhesive holds up to 80 °C.
3
Press the overhead light adhesive mount firmly onto the light.
4
Unscrew the loupe attachment with the Torx T6 screwdriver, screw on the ball-joint attachment, and keep the loupe attachment in the case.
5
Slide the ball-joint clip into the mount. Align the camera with the light spot, and tighten the joint with the screwdriver if it moves.
6
Use the white 5 m camera cable, adding the active extension cables if the computer is further away. Hold the cables in place with the clips.
Never let the camera or its cables touch the lamp's hot parts or sit in its beam.
Video: mounting on the treatment light (2019, earlier camera).
proCam XS
1
proCam XS comes with the loupe attachment installed.
2
Slide the camera onto the universal loupe clip, a brand-specific adapter for your loupes, or the Futudent headband. Tighten the attachment with the screwdriver if it moves.
3
Slide the spiral cable guide onto the loupe arm and put the cable inside. Clip the cable to your coat with the clothing clip so it doesn't pull as you work.
4
Use the black 1.5 m cable, and the active extension cables for a longer run.
Video: mounting on loupes (2019, earlier camera).
smartCam and proCam XS
1
Screw the flexible arm onto the C-clamp, then clamp it to the table and tighten.
2
Check that the ball-joint clip is on the camera, and slide it into the arm's mount. Tighten the joint with the screwdriver if it moves.
3
Use the black 1.5 m cable, or the 5 m cable or the extension cables if the computer is further away. smartCam may need a closer focus range on an arm: see Focusing.
scopeCam
scopeCam attaches to the microscope's beam splitter through a C-mount adapter made for your microscope. Mounting and adjustment are described in the scopeCam instructions for use.
Focus once, at your working distance, after mounting. Use the focus card from the camera case.
1
Put the camera in its working position: on your loupes, the headband or the light.
2
Hold the focus card from the camera case at your normal working distance.
3
Aim the camera at the centre of the card, where the lines cross, and check it on the screen. With loupes, look at the centre of the card through the loupes while you aim.
4
Watching the screen, turn the lens left or right until the thin lines are as sharp as they get.
Don't adjust the lens while you treat a patient, and don't turn it too far: the lens unscrews completely.
Video: aligning and focusing on loupes (2019, earlier camera).
smartCam focuses by itself between 50 and 120 cm, which covers a camera on the treatment light. For a closer range, for example on a flexible arm, turn the lens sleeve from its default position, where the dots on the lens and the housing line up. The autofocus then works in the new range.
Don't press against the hard stop. Otherwise leave the lens alone: turning it shifts the autofocus range and can make the image soft.
scopeCam: see the scopeCam instructions for use.
Record, take photos and set bookmarks without touching the computer.
On some Windows laptops, hold Fn with the function key: Fn + F6. Space works too.
The camera has no microphone. VISION can record sound from the computer's microphone: in Capture, open Options, then Audio, and click Record audio. It is off until you switch it on. Any connected microphone works.
At the automatic quality, an hour of 4K video is about 12 GB and an hour of Full HD about 5.4 GB. A photo is 8 megapixels.
Record a short test before treatment to check colour and light.
Wipe the camera with a soft, lint-free cloth. Disinfect with a low-alcohol or alcohol-free medical disinfectant on a wrung-out wipe. Clean the lens only with camera lens cleaner.
Don't spray the camera, submerge it or put it in an autoclave. It is not made to be sterilised.
Check that the cables are not worn and replace any with damaged insulation. Check the mounts are firm, and tighten the joints with the screwdriver.
Install the operating system updates and the latest version of VISION.
If the camera gets hotter than 70 °C, unplug it and let it cool for an hour. If it heats up again within minutes, stop using it and contact support.
Keep the camera between 5 and 50 °C, and let it reach room temperature before use so the lens doesn't fog.
The full instructions for use and the service forms, to download.
PDF, English, October 2026
proCam XS instructions for use
PDF, English, October 2026
PDF, English, June 2023
VISION 1.1 for Windows instructions
PDF, English, October 2026. Also in Spanish, French, Italian and Chinese.
PDF, English, October 2026
Extended warranty registration
PDF form
PDF form
PDF form
Set up, learn and update your camera and VISION.
Mount and connect your camera, and install VISION.
Installation guides for lights, arms and loupes, focusing, and requirements.
Short videos on setting up and using the camera.
Answers to the questions we hear most.
What changed in each version of the software.
VISION 1.2 for Windows and Mac, free for camera owners.
Register your camera with us so we can support you.
Online training for clinics, schools and distributors.